Review Combined optical imaging and mammography of the healthy breast

Paper Review

1.Combined Optical Imaging and Mammography of the Healthy Breast: Optical Contrast Derived From Breast Structure and Compression

IEEE Trans Med Imaging. 2009 January ; 28(1): 30–42. doi:10.1109/TMI.2008.925082

key words: Breast imaging; multimodality imaging; tomography

Abstract

  1. the instrument and software platform of a combined X-ray mammography/diffuse optical breast imaging system
  2. focus on system validation
  3. finite-element method for forward modeling and regularized Gauss-Newton method for parameter reconstruction
  4. enhanced coupling coefficient estimation scheme ->improve accuracy and robustness
  5. recovered average total HbT and $SO_2 $ :16.2um and 71%. HbT present a linear trend with breast density. Low HbT value may due to mammographic compression

Advancements

  1. a combined X-ray and optical breast imaging system
  2. The blood-pressure coupling effect

Methods

A block diagram of the tomographic optical imaging system(TOBI)

RF modulated imaging system and the CW system

1563503336102

TOBI optical probes (provide the capability of co-registration with 2-D mammography or tomosynthesis)

Two protocols
similarity:

a solid calibration phantom measurement is required

repeated measurements

difference:

temporal duration between the initial breast compression and the optical data acquisition

duration in protocol 2 is roughly 1 or 2 min longer than in protocol 1

expect more hemodynamic change during the measurement period in protocol 1 than in protocol 2. (why?)

Data Analysis procedure

1563505644412

image reconstruction steps:

  1. estimate the bulk optical properties of the patient’s breast and the mean source/detector coupling coefficients
  2. perform a full image reconstruction starting with the homogeneous initial guess from the output of the previous step
image reconstruction algorithm

iterative Gauss-Newton reconstruction approach

Forward Model

diffusion equation $\mu_a$ :absorption coefficient

$D(r)=\frac{1}{3}(\mu^{’}_a(r)+\mu_s(r)) \approx \frac{1}{3} \mu^{’}_s(r)$ :diffusion coefficient

$\mu^{’}_s$ :reduced scattering coefficient

$c=c_0/n $ :speed of light in the medium

$S_0(r,t)$ :light source

frequency-domain diffusion equation discretization: $$

\sum_{i} \Phi_{i} \sum_{k}\left(D_{k}\left\langle\phi_{k}, \nabla \varphi_{i} \cdot \nabla \varphi_{j}\right\rangle- D_{k}\left\langle\phi_{k} \nabla \varphi_{i}, \varphi_{j}\right\rangle_{\partial \Omega}+\left(\mu_{a}^{k}+\frac{j \omega}{c}\right)\left\langle\phi_{k} \varphi_{i}, \varphi_{j}\right\rangle\right)=\left\langle S_{0}, \varphi_{j}\right\rangle $$ $\phi_k$ :basis on the reconstruction mesh

$\varphi_i$ and $\varphi_j$ :basis and weight functions on the forward mesh

$\langle f,g \rangle$ : intergration $\int_{\Omega} fg dr$ over volume $\Omega$

$\langle\vec{f}, g\rangle_{\partial \Omega}$ : surface integration $\int_{\partial \Omega} \vec{g} \vec{f} \cdot d \vec{S}$ on the boundaries $\partial \Omega$

->final form for the forward quation(6): linear equation in form of $A \mathbf{x}=\mathbf{b}$

Mesh Generation
  1. create a uniform 3-D grid
  2. split each cube based on a T5 rule

dual-mesh scheme: a forward mesh for diffusion modeling and a separate reconstruction mesh to represent the optical properties

different mesh densities, the higher one for the forward mesh, the lower one for the reconstruction mesh

Inverse Problem

nonlinear optimization problem (7) Gauss-Newton method updates $\mu_a$ and $D$ vectors iteratively (8) $\triangle_{K} \mu_{a}$ and $\Delta_{k} D$ : property updates of absorption and diffusion coefficients

$\gamma_{k}$ : determined by empirical approach

$J_k$ : Jacobian matrix defined as $J_{k}=\left(J_{\mu_{a}} \quad J_{D}\right)$ (9) a heuristic re-weighting scheme when solving (8):

Data Calibration and Source/Detector Coupling Coefficient Estimation

(16) $\Phi_{T}^{\text { meas }}, \Phi_{p}^{\text { meas }},$ and $\Phi_{p}^{\text { model }}$ : measured target data, measured and model prediction of calibration phantom data

not sufficient to remove all the systematic errors

simultaneous determination of coupling coefficients (18)

advantages: greatly reduce the previously mentioned artifacts

side effect: reduce target contrast

(24) perform high-pass filter on SD distributions

Bulk Property Estimation and Image Reconstruction

estimated optical properties were used as homogeneous initial guess for the image reconstruction

can not be used for the second stage (SD coefficients maximum coupling to the optical images)

Results

Phantom reconstruction

maximum absorption contrast are lower than expected contrast 2:1

Clinical Data Reconstructions

1563589977757

For breasts with very little glandular content, the averaged HbT values are relatively small

For breasts with small amount of glandular tissue, a positive contrast in HbT from the bulk values at the glandular region appears in the images

pressure simulations

expect lower-HbT content in the region where tissue stress is stronger

breast tumors have been shown to have even higher stiffness and interstitial pressure than normal breast tissue

The additional complexities of the tissue compression response represent an opportunity for identifying dynamic characteristics useful in cancer diagnosis.

Problems and further study
  1. expect more hemodynamic change during the measurement period in protocol 1 than in protocol 2. (why?)

    (temporal duration between the initial breast compression and the optical data acquisition duration in protocol 2 is roughly 1 or 2 min longer than in protocol 1, why more duration caused less change?)

  2. tissue compression response ->dynamic characteristics for cancer diagnosis

  3. only healthy breasts in this paper->further study in tumor cases